HIOKI
  • Testing voltage 0.2 V AC to 5.00 kV AC Voltage setting Digital setting, Setting resolution: 0.01 kV Waveform/Frequency Sine wave (Distortion ratio 5 % or less at no load), 50/60 Hz selectable Current measurement 0.01 mA to 20.0 mA, True RMS rectified (digital display) Measurement range 10 mA (0.01 mA resolution), 20 mA (0.1 mA resolution) Voltage meter Accuracy: ±1.5 % rdg. (1000 V or more), ±15 V (less than 1000 V), True RMS rectified Judgment function Window comparator method (Digital setting) [Insulation test section] Testing voltage 500 V DC, 1000 V DC Unloaded voltage 1 to 1.2 times rated voltage Rated testing current 1 to 1.2 mA, Shorted current: 4 to 5 mA (at 500 V), 2 to 3 mA (at 1000 V) Measurement range, Accuracy 0.5 MΩ to 999 MΩ (at 500 V), and 1 MΩ to 999 MΩ (at 1000 V): ±4 % rdg., 1000 MΩ to 2000 MΩ: ±8 % rdg. Judgment function Window comparator method (Digital setting) [Timer section] *Test times may differ from set timer times depending on the load. Setting range 0.3 to 999 s Ramp, Delay Testing voltage ramp-up, or down, Insulation test delay: 0.1 to 99.9 s [General section] Functions Saving 8 testing conditions, hold, buzzer, contact check Monitor function Output voltage, detected current, insulation resistance, Refresh rate: 2 times/s Power supply 100 to 240 V AC, (50/60 Hz), 200 VA max.
Số lượng
Gọi tư vấn
Testing voltage 0.2 V AC to 5.00 kV AC
Voltage setting Digital setting, Setting resolution: 0.01 kV
Waveform/Frequency Sine wave (Distortion ratio 5 % or less at no load), 50/60 Hz selectable
Current measurement 0.01 mA to 20.0 mA, True RMS rectified (digital display)
Measurement range 10 mA (0.01 mA resolution), 20 mA (0.1 mA resolution)
Voltage meter Accuracy: ±1.5 % rdg. (1000 V or more), ±15 V (less than 1000 V), True RMS rectified
Judgment function Window comparator method (Digital setting)
[Insulation test section]
Testing voltage 500 V DC, 1000 V DC
Unloaded voltage 1 to 1.2 times rated voltage
Rated testing current 1 to 1.2 mA, Shorted current: 4 to 5 mA (at 500 V), 2 to 3 mA (at 1000 V)
Measurement range, Accuracy 0.5 MΩ to 999 MΩ (at 500 V), and 1 MΩ to 999 MΩ (at 1000 V): ±4 % rdg., 1000 MΩ to 2000 MΩ: ±8 % rdg.
Judgment function Window comparator method (Digital setting)
[Timer section] *Test times may differ from set timer times depending on the load.
Setting range 0.3 to 999 s
Ramp, Delay Testing voltage ramp-up, or down, Insulation test delay: 0.1 to 99.9 s
[General section]
Functions Saving 8 testing conditions, hold, buzzer, contact check
Monitor function Output voltage, detected current, insulation resistance, Refresh rate: 2 times/s
Power supply 100 to 240 V AC, (50/60 Hz), 200 VA max.